Open Access System for Information Sharing

Login Library

 

Article
Cited 20 time in webofscience Cited 20 time in scopus
Metadata Downloads
Full metadata record
Files in This Item:
There are no files associated with this item.
DC FieldValueLanguage
dc.contributor.authorYang Joon Kang-
dc.contributor.authorEun Seop Yeom-
dc.contributor.authorLee, SJ-
dc.date.accessioned2016-03-31T08:06:56Z-
dc.date.available2016-03-31T08:06:56Z-
dc.date.created2014-03-20-
dc.date.issued2013-11-05-
dc.identifier.issn0003-2700-
dc.identifier.other2013-OAK-0000029698-
dc.identifier.urihttps://oasis.postech.ac.kr/handle/2014.oak/14594-
dc.description.abstractIn this study, we propose a novel microfluidic biosensor for monitoring hemorheological and hemodynamic properties using an extracorporeal rat bypass loop. To monitor temporal variations of biophysical properties including viscosity, flow rate, and pressure of rat blood, a complex fluidic network is established by connecting the abdominal aorta and jugular vein to an extracorporeal bypass loop including a flow stabilizer and a microfluidic biosensor. Three biophysical properties are simultaneously measured through label-free operation and sensorless detection based on two sequential flow controls in the microfluidic channel. A discrete fluidic-circuit model is employed to derive analytical formulas for the complex fluidic network. First, to evaluate the measurement accuracy of the proposed method, a peristaltic pump is used as substitute for a rat. The flow rate and viscosity of 20% glycerin (test fluid) circulating within the fluidic network are measured, and then the results are compared with those obtained using conventional methods. The normal differences between two measurement methods are less than 4%. Then, the proposed method is used to monitor temporal variations in biophysical properties of blood circulating within the complex fluidic network under normal and continuous conditions. Rats require at least 30 mm to adapt to different fluidic environments. The flow rate, pressure, and hematocrit of rat blood tend to decrease gradually because of continuous hemodilution effect. Furthermore, the reduced flow rate increases blood viscosity under hemodilution condition. These experiments demonstrate that the proposed method can effectively monitor temporal variations of biophysical properties of rat blood under ex vivo conditions.-
dc.description.statementofresponsibilityX-
dc.languageEnglish-
dc.publisherAMER CHEMICAL SOC-
dc.relation.isPartOfAnalytical chemistry-
dc.subjectCELL-FREE LAYER-
dc.subjectIN-VIVO-
dc.subjectBLOOD-
dc.subjectVISCOSITY-
dc.subjectAGGREGATION-
dc.subjectVISCOMETER-
dc.subjectFLUID-
dc.subjectDEFORMABILITY-
dc.subjectRHEOLOGY-
dc.titleMicrofluidic biosensor for monitoring temporal variations of hemorheological and hemodynamic properties using an extracorporeal rat bypass loop-
dc.typeArticle-
dc.contributor.college기계공학과-
dc.identifier.doi10.1021/AC402505Z-
dc.author.googleKang, YJ-
dc.author.googleYeom, E-
dc.author.googleLee, SJ-
dc.relation.volume85-
dc.relation.issue21-
dc.relation.startpage10503-
dc.relation.lastpage10511-
dc.contributor.id10054593-
dc.relation.journalAnalytical chemistry-
dc.relation.indexSCI급, SCOPUS 등재논문-
dc.relation.sciSCI-
dc.collections.nameJournal Papers-
dc.type.rimsART-
dc.identifier.bibliographicCitationAnalytical chemistry, v.85, no.21, pp.10503 - 10511-
dc.identifier.wosid000326711400074-
dc.date.tcdate2019-01-01-
dc.citation.endPage10511-
dc.citation.number21-
dc.citation.startPage10503-
dc.citation.titleAnalytical chemistry-
dc.citation.volume85-
dc.contributor.affiliatedAuthorLee, SJ-
dc.identifier.scopusid2-s2.0-84887734891-
dc.description.journalClass1-
dc.description.journalClass1-
dc.description.wostc11-
dc.description.scptc5*
dc.date.scptcdate2018-05-121*
dc.type.docTypeArticle-
dc.subject.keywordPlusCELL-FREE LAYER-
dc.subject.keywordPlusIN-VIVO-
dc.subject.keywordPlusBLOOD-
dc.subject.keywordPlusVISCOSITY-
dc.subject.keywordPlusAGGREGATION-
dc.subject.keywordPlusVISCOMETER-
dc.relation.journalWebOfScienceCategoryChemistry, Analytical-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaChemistry-

qr_code

  • mendeley

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher

이상준LEE, SANG JOON
Dept of Mechanical Enginrg
Read more

Views & Downloads

Browse