Open Access System for Information Sharing

Login Library

 

Article
Cited 7 time in webofscience Cited 8 time in scopus
Metadata Downloads

Effect of the Cr-rich oxide surface on fast pumpdown to ultrahigh vacuum SCIE SCOPUS

Title
Effect of the Cr-rich oxide surface on fast pumpdown to ultrahigh vacuum
Authors
Chun, ICho, BChung, S
Date Issued
1997-09
Publisher
AMER INST PHYSICS
Abstract
The outgassing characteristic of an ordinary stainless-steel ultrahigh vacuum (UHV) chamber was determined using an extremely dry nitrogen venting system, and compared with that of an extreme high vacuum (XHV) chamber. A marked sorption resistance effect of the vacuum surface of the XHV chamber was found. The outgassing rate of the XHV chamber was about 100 times lower than that of the ordinary UHV chamber in the unbaked case. It is believed that the recent achievement of fast pumpdown to UHV is the result of not only the reduction of the water content of the venting gas, but also the low gas-sorability of the vacuum surface. (C) 1997 American Vacuum Society.
Keywords
EXTREME HIGH-VACUUM; STAINLESS-STEEL; ALUMINUM; SYSTEM
URI
https://oasis.postech.ac.kr/handle/2014.oak/21219
DOI
10.1116/1.580763
ISSN
0734-2101
Article Type
Article
Citation
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, vol. 15, no. 5, page. 2518 - 2520, 1997-09